Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.

¾È°æ¿ø¿¡¼­ È°¿ë °¡´ÉÇÑ Å丯¼ÒÇÁÆ®·»Áî ȸÀüÆò°¡ ¹æ¹ýÀÇ À¯È¿¼º Æò°¡ - ÈÞ´ë¿ë µ¸º¸±â, °¢µµ±â ¾îÇø®ÄÉÀ̼Ç, ÇÚµåÆùÀÇ »çÁø ÃÔ¿µ Evaluation of the Effectiveness of the Toric Soft Lens¡¯s Orientation-Evaluate Method Available at the Optometry Practice

´ëÇѽðúÇÐȸÁö 2020³â 22±Ç 4È£ p.375 ~ 385
Á¶ÇÏ¿ø, Á¤¿¹¸°, ¹ÚÇü¹Î, Ãߺ´¼±,
¼Ò¼Ó »ó¼¼Á¤º¸
Á¶ÇÏ¿ø ( Jo Ha-Won ) 
Daegu Catholic University Department of Optometry and Vision Science

Á¤¿¹¸° ( Jung Yee-Rin ) 
Daegu Catholic University Department of Optometry and Vision Science
¹ÚÇü¹Î ( Park Hyung-Min ) 
Daegu Catholic University Department of Optometry and Vision Science
Ãߺ´¼± ( Chu Byoung-Sun ) 
Daegu Catholic University Department of Optometry and Vision Science

Abstract


Purpose : The purpose of this study was to evaluate the agreement of different methods on assessing toric lens rotation.

Methods : After fitting toric soft lens according to the manufacturer's guidelines at model eye and 57 eyes (average 24.79¡¾1.29 years old), rotational evaluation was performed using a mobile phone photography, protractor application, portable magnifier and slit-lamp. Agreements among different methods and two testers were analyzed using Bland-Altman analysed.

Results : When the rotation amount of the toric soft lens was measured by the same rotation evaluation method with different testers, the average difference of the measured values was ?0.50, -0.70, -0.30 and ?0.13¡Æ showing similar results. Regarding the comparison of orientationevaluate methods, when using the mobile phone photography and protractor App, the measured values within the tolerance range were relatively low with 74.56% and 75.44%, respectively, and with the portable magnifier, the reliability was high with 95.61%.

Conclusion : Even if a slit-lamp can not be used in the prescription of a toric soft lens, clinically useful results for toric soft lens rotation evaluation can be obtained through photography, mobile phone protractor application and portable magnifier.

Å°¿öµå

LARS; Orientation evaluation; Portable magnifier; Rotation amount; Toric soft contact lens

¿ø¹® ¹× ¸µÅ©¾Æ¿ô Á¤º¸

µîÀçÀú³Î Á¤º¸

KCI